Influence of Binarization Process on Vascular Density Metrics: A Quantitative Optical Coherence Tomography Angiography Assessment in Human and Porcine Retinas

Background Optical Coherence Tomography Angiography (OCTA) provides high-resolution visualization of retinal microvasculature, with vascular density (VD) serving as a one of the key quantitative metrics. However, VD measurements are highly sensitive to image binarization step, and no standardized approach exists.

Methods We analyzed 51 OCTA scans (human and porcine) using 29 binarization algorithms, including global and local thresholding techniques from ImageJ and DoxaPy, as well as Random Walker segmentation. VD was calculated for each binarization algorithm and compared against Optovue-generated values (ground truth). Results were evaluated using hierarchical clustering and agreement between them was determined by Bland–Altman analysis.

Results Wolf algorithm was found to exhibit least deviation from mean Optovue VD values for human SCP layer (46.5 ± 1.2% vs. 48.3 ± 1.4%; p = 3.62 x 10-5); however, there is not significant difference between VDs from Optovue and Wolf algorithms from porcine SCP layer (46.2 ± 1.8 % vs 46.3 ± 1.4% ; p=0.74). For DCP layer, Phansalkar algorithm exhibited least VD variability (50.7 ± 2.0% vs. 51.9 ± 1.7%; p=2.53 x 10-4) in the human cohort. Whereas Percentile algorithm exhibited least, non-significant variations in the porcine DCP layer VD (50.0 ± 1.4 % vs. 50.3 ± 1.4%; p=0.75).

Discussion Each binarization technique evaluated in this study impacts OCTA-derived VD measurements differently. Local adaptive algorithms collectively outperform global methods, particularly for SCP analysis. Standardization of image processing pipelines and layer-specific optimization are essential to improve reproducibility and clinical consistency.

Competing Interest Statement

The authors have declared no competing interest.

Funding Statement

Research reported in this publication was supported by the National Institutes of Health under award number 1R21NS135307-01 and the Department of Anesthesiology and Pain Management of the University of Texas Southwestern Medical Center. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health or the University of Texas Southwestern Medical Center.

Author Declarations

I confirm all relevant ethical guidelines have been followed, and any necessary IRB and/or ethics committee approvals have been obtained.

Yes

The details of the IRB/oversight body that provided approval or exemption for the research described are given below:

IRB Committee of The University of Texas Southwestern Medical Center gave ethical approval for this work.

I confirm that all necessary patient/participant consent has been obtained and the appropriate institutional forms have been archived, and that any patient/participant/sample identifiers included were not known to anyone (e.g., hospital staff, patients or participants themselves) outside the research group so cannot be used to identify individuals.

Yes

I understand that all clinical trials and any other prospective interventional studies must be registered with an ICMJE-approved registry, such as ClinicalTrials.gov. I confirm that any such study reported in the manuscript has been registered and the trial registration ID is provided (note: if posting a prospective study registered retrospectively, please provide a statement in the trial ID field explaining why the study was not registered in advance).

Yes

I have followed all appropriate research reporting guidelines, such as any relevant EQUATOR Network research reporting checklist(s) and other pertinent material, if applicable.

Yes