Chronic venous occlusions (CVOs) occur when veins become obstructed due to thrombus formation, with subsequent organization into an occlusion by collagenous material.[1] [2] Iliofemoral venous obstruction is a well-recognized cause of lower extremity venous hypertension and may manifest as limb swelling, pain, venous claudication, skin changes, or ulceration in advanced disease.[1] [3] Endovascular therapy has become the primary treatment strategy for symptomatic iliofemoral venous obstruction, with the goal of restoring venous outflow and improving quality of life. In many cases, this requires venous stent placement to address fixed anatomic obstruction and maintain long-term patency. Despite technical advances, venous stents are vulnerable to failure over time, most commonly due to in-stent restenosis (ISR) or stent occlusion.
The primary patency rates of venous stents depend on multiple factors, including the inflow-outflow veins, adherence to anticoagulation therapy, and mechanical failure of the stent.[4] Maintaining long-term stent patency remains a significant challenge, as patency rates decline within 4 to 5 years in lower extremity vessels, often resulting in recurrence of symptoms and reintervention. ISR, stent compression, or total occlusion of the stents is associated with recurrence of the symptoms.[4] Restenosis and occlusion of venous stents are primarily characterized by thrombosis and subsequent organization into type 1 and 3 collagen.[2] [5] [6] [7] [8] Stents placed for postthrombotic obstructions have primary patencies reported as 88% at 12 months, and 81.6% at 36 months.[9] [10] Projected 5-year primary patency in chronic postthrombotic obstructions is approximately 60%.[11]
Crossing stent occlusions represents a rate-limiting step in endovascular intervention. Initial attempts to recanalize venous stent occlusions involve conventional tools, for example, hydrophilic guidewires advanced through well-supported coaxial or triaxial catheter systems to penetrate the occlusion.[3] The success of conventional tools relies on the physical characteristics of the occlusion. The composition of chronic stent occlusions is variable and classified into two types: soft lesions and hard lesions.[6] [12] [13] Softer lesions are often thrombotic in nature, whereas more resistant hard lesions consist of organized thrombus with dense collagen deposition, fibrosis, or calcification. These rigid, hard occlusions require substantially greater penetration force than soft thrombotic lesions and frequently challenge conventional guidewire techniques. Failure to cross may result from wire buckling, loss of directional control, or deviation from the intended intraluminal course.[3] When conventional tools fail to cross occlusions, several sharp needle-based crossing tools can be deployed.[1] [14] [15] [16] [17] [18] [19] However, there remains a subset of cases with heavily fibrotic occlusions that are resistant to both conventional and sharp recanalization tools and require alternative tools for crossing the occlusion.
The PowerWire (Baylis Medical Technologies, Toronto, Canada) radiofrequency (RF) guidewire-assisted recanalization has shown high success rates in crossing occlusions in both native and stented vessels.[3] [13] [20] [21] [22] [23] [24] [25] [26] RF-assisted recanalization has been applied across a range of complex refractory vascular occlusions, including peripheral arterial chronic total occlusions, central venous occlusions, and advanced iliocaval disease. This technology enables controlled delivery of RF energy at the wire tip, facilitating advancement through dense fibrotic lesions with less mechanical force than required for conventional guidewires.[27] By reducing the reliance on forward pressure alone, the RF-wire assisted technique may mitigate failure to cross due to buckling or uncontrolled deflection. Additionally, an advantage of RF-wire during stent recanalization is its response to metal contact.[3] Upon contact with the stent material, the system automatically interrupts RF energy delivery, limiting further advancement and potentially reducing the risk of unintended passage beyond the stent lumen. This attribute may be particularly useful when navigating long or complex stent occlusions, where maintaining a central intraluminal trajectory is critical.
The purpose of this article is to describe our experience in utilizing RF wire to cross stent occlusions within the iliofemoral venous system. We present a series of cases ([Figs. 1] [2] [3] [4]) where PowerWire RF wire was successfully used to cross complex stent occlusions. We also report a case in which a chronically occluded left common/external iliac stent appears to have eroded into the adjacent extravascular space, no longer sharing the common channel with the native common iliac vein and infrarenal IVC ([Fig. 4]). This case emphasizes the importance of comprehensive preprocedural planning using cross-sectional imaging to accurately delineate vascular anatomy and identify adjacent structures at risk for injury. Furthermore, it highlights the need for employing multiple intraprocedural fluoroscopic obliquities and meticulous triangulation to ensure an appropriate intraluminal trajectory is maintained.